PURELAB CHORUS 3 RO 20L/HR WITHOUT BOOST PUMP
PURELAB CHORUS 3 RO 20L/HR WITHOUT BOOST PUMP PURELAB CHORUS 3 RO 20L/HR WITHOUT BOOST PUMP

PURELAB CHORUS 3 RO 20L/HR WITHOUT BOOST PUMP

Part Number: RO320XXM2

Ships within: 4 weeks

Quick Overview

The PURELAB Chorus gives you the chance to configure a solution that best suits your needs with the added benefit of flexibility should your future needs change. Ideal for a wide range of applications including sophisticated research and testing applications, qualitative analyses, feeding ultrapure water systems, glassware washing, autoclaves etc. The list of possible applications is endless!


Featuring:


PURELAB Chorus 1: ultrapure water for critical applications


PURELAB Chorus 2: for your pure Type II+/II solutions


PURELAB Chorus 3: for your general grade Type III solutions Halo Dispense Solutions: three remote dispensing and monitoring solutions


Storage Reservoirs: designed to accomodate PURELAB Chorus water purification systems

  • Details
  • Specifications
  • Feed Water
  • Installation
  • Dimensions & Weights
Local SKU RO320XXM2
Main Feature

The PURELAB Chorus gives you the chance to configure a solution that best suits your needs with the added benefit of flexibility should your future needs change. Ideal for a wide range of applications including sophisticated research and testing applications, qualitative analyses, feeding ultrapure water systems, glassware washing, autoclaves etc. The list of possible applications is endless!

Featuring:

PURELAB Chorus 1: ultrapure water for critical applications

PURELAB Chorus 2: for your pure Type II+/II solutions

PURELAB Chorus 3: for your general grade Type III solutions Halo Dispense Solutions: three remote dispensing and monitoring solutions

Storage Reservoirs: designed to accomodate PURELAB Chorus water purification systems

Brand Name Range PURELAB
Validation Support Available Yes
Applications Feed Autoclave / steriliser, Feed to ultra pure water systems, Glassware washing/rinsing, Heating system filling, Hydroponics, Plant growth cabinets, Stability chambers, Steam generators, Steriliser feed
Daily Usage (8hr day) > 100 l/day
Technologies Pre-treatment, Reverse Osmosis
Inorganics > 95% rejection
Organics (TOC) < 50 ppb
Bacteria < 5 CFU/ml
Endotoxins N/A
Particles >99% rejection
Footnotes 1 Standard conditions are 4 bar inlet pressure at 15 degrees centigrade, fed with potable water and a clean pre-treatment cartridge. Refer to flow tables outside these conditions. 2 Subject to correct operating and maintenance procedures
Feedwater Nature Potable supply (Tap Water)
Conductivity <2000 µS/cm
Total Chlorine 0.5 ppm
Free Chlorine <0.05 ppm
Carbon Dioxide <0.1 ppm
Silica <2 ppm
Temperature 1-35 °C
Drain (gravity fall with air gap) 80 l/hr
Min. Pressure 4 bar (60 psi) without internal boost pump
Max. Pressure 6 bar (90 psi) without internal boost pump
Footnotes Deionization cartridge life may vary with feedwaters >1400 μS/cm 4 Fit LA652 Regulator where feedwater pressure exceeds specified limits.
Installation Options Bench
Mains Input 100-240V ac, 50-60 Hz
System Voltage 24V DC
Power Consumption 60 VA (without boost pump)
Noise Level <45 dBA
Height (mm(inch)) 900 (35.4)
Width (mm(inch)) 236 (9.3)
Depth (mm(inch)) 374 (14.7)
Operating Weight (kg(lb)) 17 (37)
Footnotes Flexible Height - 900 to 1020 mm